Need help? Spain: 93 674 88 00 Int. Calls +34 93 669 86 50
AREA/LANGUAGE
Choose your country or area Language: Spanish/English
America Europe Africa Asia Oceania

Pico

  • Safety: the generator can work unattended and independently 24 hours a day, 365 days a year without requiring surveillance or programming
  • Automatic stand-by: if the production flow exceeds the consumption, the generator enters a stand-by mode to reduce energy consumption
  • Oxygen analyser: included in every N2 FLO, PICO & MAXI generators, to keep the nitrogen purity under control in real time
  • Custom nitrogen purity: the nitrogen purity can be set according to customer and application requirements

 

More info

Specifications
Downloads

MODELS

Outgoing nitrogen flow* [Nm3/h]

Purity

99,999%

99,99%

99,9%

99,5%

99%

98%

97%

N2 PICO3

1,5

2,5

5,4

8,6

10,4

13,8

16,1

N2 PICO4

2

3,3

6,9

11,5

15

18,4

20.7

N2 PICO5

2,3

4,0

9,2

14,4

18,4

23

26,5

N2 PICO6

3,5

5,0

10,4

17,3

21,9

27,9

31,1

Air/N2 Ratio

10,0

7,0

4,3

3,3

3,1

2,9

2,7

 * The flow capacity refers to inlet air nominal conditions: pressure of 8,5 bar, CMS temperature of 20 °C. If the inlet air pressure is different, apply the correction factors reported below.

 

TECHNOLOGIES EMPLOY PSA CLAIND
INLET AIR SPECIFICATIONS Air pressure : min 6,5 / max 13 bar
Quality: dry, clean, exhalation free. The inlet air must meet the specifications ISO 8573-1, class quality 1.4.1.
NITROGEN OUTLET PRESSURE Given the pressure P of the inlet air, the pressure of the nitrogen generated is 1 ÷ 2.5 bar lower than P, depending on the flow rate of nitrogen supplied
MOISTURE CONTENT AT OUTLET -50°C Dew Point ATP
ELETTRICAL SUPPLY 230 Vac ± 10% 50 Hz
115 Vac ± 10% 60 Hz
POWER RATING 50 W
INDEX OF PROTECTION IP20
AVERAGE NOISE ≤ 60 dB (measured at 1 meter from the generator)
OPERATING TEMPERATURE 5 ÷ 40°C
DIMENSIONS Height: 138 cm
Width: 40 cm
Depth: 130 cm
WEIGHT PICO 3: 230 kg
PICO 4: 270 kg
PICO 5: 310 kg
PICO 6: 350 kg
Choose your product
showing 2 of 2 products
Reference CL-446.04.0300
Description Generador PICO 6, 99,9 % purez a, con analizador  de oxígeno y tanque buffer externo de 720 litros
To use the Favorites List you must be registered Sign up
Reference CL-448.04.0220
Description Generador N2 PICO6, 99,9% 230 V / 50 Hz
To use the Favorites List you must be registered Sign up
showing 2 of 2 products